Feature Tower Mount Desktop Mount
Average Accuracy 1 down to 0.15° (0.25° to 0.5° typical) Same as Tower Mount
Sampling rate 2
EyeLink 1000: 1000 Hz monocular
EyeLink 2K: 2000 Hz monocular
EyeLink 1000: 1000 Hz mono / 500 Hz binoc
EyeLink 2K: 2000 Hz mono / 1000 Hz binoc
End-to-End Sample Delay
Monocular3
< 1.4 msec @ 2000 Hz; 1.8 msec @ 1000 Hz Same as Tower Mount
Binocular3
Not Available < 2.6 msec @ 1000 Hz; 4.8 msec @ 500 Hz.
Spatial Resolution 4 < 0.01° RMS @ 1000 Hz, 0.02° RMS @ 2000 Hz Same as Tower Mount
Eye Tracking Principal 5
Dark Pupil - Corneal Reflection Same as Tower Mount
Pupil Detection Algorithms
Centroid or Ellipse Fitting Same as Tower Mount
Pupil Size Resolution 0.2% of diameter Same as Tower Mount
Gaze Tracking Range 55° Horizontal, 45° Vertical
32° Horizontal, 25° Vertical
Allowed Head Movement 6 50 mm Horizontal and Vertical, 20 mm Depth
Same as Tower Mount
Glasses Compatibility
Good Excellent
Online Event Parsing Fixation / Saccade / Blink / Fixation Update Same as Tower Mount
Realtime Operator Feedback Eye position cursor or position traces. Camera images and tracking status. Same as Tower Mount
Eye - Camera distance
Fixed at about 38 cm
Between 40 - 70 cm
Eye Illumination
910 nm, Class 1 LED Product, <1.0 mW/cm2 in standard configuration
890 nm, Class 1 LED Product, <1.0 mW/cm2 in standard configuration
Digital Data Access Ethernet Ethernet
Analog Output Optional PCI card Optional PCI card
Response Box 7 Button USB response pad included 7 Button USB response pad included
Host Operating System DOS- RTOS DOS- RTOS
Display Operating System Windows, Linux, Mac OSX, Mac OS9, DOS Windows, Linux, Mac OSX, Mac OS9, DOS
Approvals FCC and CISPR Class A, 60950-1 ITE Equipment Same as Tower Mount
 
(1) Measured with real eye fixations at multiple screen positions
(2) Native sampling rate. 250 Hz and 500 Hz downsampling available
(3) Measured from an actual physical event to availability of first data sample
that registered the event on the Display / Subject PC via Ethernet or Analog output.
If enabled, host PC filtering adds one sample extra delay for each filtering level."
(4) Measured with an artificial pupil
(5) Pupil Only tracking mode is available for use in head fixed conditions
(6) Binocular tracking with desktop mount can reduce allowed head movement
to approx. 25 mm Horizontal and Vertical




 